Abstract: Particle morphology can influence the particle-size distributions obtained by different measurement techniques. This study compares mechanical sieving (MS), static light scattering (SLS), and dynamic image analysis (DIA) for a bimodal zirconia sphere mixture, sea sand, and 3D-printed pyramids and cylinders.
All methods resolved the two sphere populations, with D50 values of approximately 630 µm for MS, 599 µm for SLS, and 487 µm for DIA using the minimum-chord descriptor. For sea sand, MS and SLS gave D50 values of 194 and 210 µm, while three DIA descriptors yielded similar medians of 200–213 µm. Larger differences were observed for the 3D-printed particles. Pyramid D50 values were approximately 1450 µm for MS, 1310 µm for SLS, and 1552–1812 µm for DIA. For cylinders, DIA medians ranged from 1026 to 3123 µm, spanning the transverse diameter and nominal particle length; SLS produced a broad distribution.
The results show that particle morphology influences the agreement between methods. Several DIA descriptors aligned with MS and SLS for sea sand, whereas the regularly anisotropic printed particles showed pronounced differences among DIA descriptors. Particle-size results should therefore be interpreted in relation to particle morphology and the measurement principle.